Health monitoring design for wearable devices

I was unable to find a phone call 20 years ago. Twenty years later, mobile phones allow us to make calls anytime, anywhere. After 20 years of innovation, voice calls are no longer the main function of this smart device, it can not only take beautiful photos, play audio and video streaming files, but also provide a variety of services - now It has also gradually become our personal trainer. When equipped with sensors or connected to wearable sensors, these devices can be used to monitor daily activities and personal health. Driven by increasing health awareness, people are beginning to focus on measuring vital signs such as heart rate, body temperature, oxygen saturation, blood pressure, activity levels (activity) and fat burning – and tracking the daily trends of these parameters.

These parameters are now monitored by a universal sensor front end with multiple sensors. The biggest challenge is to minimize size and extend battery life. This article discusses solutions for the rapidly growing market for wearable electronics.

Most important vital signs

Without heartbeat, we will have big troubles, so the pulse or heart rate is still the most important parameter we need to monitor. In addition to the number of beats per minute, we also want to check the relationship between cardiac behavior and activity. Heart rhythm is also very important because the rapidly changing heart rate is a sign of heart disease.

Heart rate and cardiac activity monitoring are typically achieved using electrocardiogram (ECG) measurements of physiological electrical signals. An electrode connected to the body measures the activity of the signal of the heart tissue of the heart tissue. Based on this principle, the professional diagnostic system can connect up to 10 electrodes in the chest and limbs. The ECG provides detailed information about the different components of the heartbeat (P wave, QRS and T waves).

Health monitoring of wearable devices

Figure 1. AD8232 Single Lead ECG Front End

The use of single-lead ECGs in the sports world is becoming more common, using a two-electrode chest strap to measure cardiac activity. Although various ECG waveforms can be detected, most systems only measure heart rate. These chest straps are uncomfortable to wear, so the sports and healthcare industry is looking for alternatives, such as integrating electrodes into sweatshirts. The AD8232 single-lead heart rate monitor front end (shown in Figure 1) was developed specifically for such low-power wearable applications. The device incorporates an instrumentation amplifier with a gain of 100 V/V and a high-pass filter that blocks the offset voltage generated by the half-cell potential of the electrodes on the skin. The output buffer and low-pass filter suppress high-frequency components (EMG signals) generated by muscle activity. This low-power front-end consumes 170μA and can be used with the 16-bit on-chip meter ADuCM350 for high-performance, single-lead ECG measurements.

A new trend in heart rate measurement is the Photoelectric Volume Chart (PPG), an optical technique that obtains cardiac function information without measuring bioelectrical signals. PPG is primarily used to measure blood oxygen saturation (SpO2), but it can also provide cardiac function information without bioelectrical signal measurements. With PPG technology, the heart rate monitor can be integrated into wearable devices such as watches or wristbands. This is not possible because the signal level of the physiological potential method is extremely weak.

In an optical system, light is projected from the surface of the skin. The amount of light absorbed by the red blood cells is measured by a photosensor. As the heart beats, the ever-changing blood volume disperses the amount of light received. When measuring on a finger or earlobe, the best accuracy is obtained with a red or infrared source due to the considerable amount of arterial blood in these areas. However, there are few arteries on the surface of the wrist, and wrist-worn devices must detect the pulsating component through the veins and capillaries below the surface of the skin, so the green light effect is better.

The ADPD142 optical module (shown in Figure 2) has a complete photometric front end with integrated photosensors, current sources and LEDs. Designed to measure reflected light, the device can be used to implement PPG measurements. All components are packaged in a small module.

Health monitoring of wearable devices

Figure 2. ADPD142 optical module

The challenge of using optical VSM

The main challenge in measuring PPG with wrist-worn devices comes from ambient light and motion-induced interference. The DC error generated by sunlight is relatively easy to eliminate, but the light emitted by fluorescent lamps and energy-saving lamps has a frequency component that causes an AC error. The analog front end uses two structures to reject interference signals from DC to 100 kHz. After the analog signal is conditioned, the 14-bit successive approximation digital-to-analog converter (ADC) digitizes the signal and sends it to the microprocessor for final post-processing via the I2C interface.

The synchronous transmit path is integrated in parallel with the optical receiver. Its independent current source drives two separate LEDs with current levels up to 250 mA. The LED current is pulsed and the pulse length is in the microsecond range, thus maintaining a low average power consumption for maximum battery life.

The LED driver circuit is dynamic and ready to configure, so it is not affected by various environmental conditions, such as ambient light, the color of the wearer's skin and hair, or the sweat between the sensor and the skin, which reduces sensitivity. The excitation LED configuration is very convenient and can be used to build an adaptive system. All timing and synchronization are handled by the analog front end, so there is no added overhead for the system processor.

The ADPD142 is available in two versions: the ADPD142RG integrates red and green LEDs to support optical heart rate monitoring; the ADPD142RI integrates red and infrared LEDs for oxygen saturation (SpO2) measurements. )

Impact of exercise

Movement can also interfere with the optical system. This may not be a problem when an optical heart rate monitor is used for sleep studies, but sports watches and wristbands will have difficulty eliminating motion artifacts if worn during exercise. The relative motion between the optical sensor (LED and photodetector) and the skin reduces the sensitivity of the optical signal. In addition, the frequency component of the motion may also be considered a heart rate measurement, so the motion must be measured and compensated. The closer the device is to the human body, the smaller the effect, but it is almost impossible to mechanically eliminate this effect.

We can use a variety of methods to measure motion. One of them is an optical method that uses multiple LED wavelengths. The common mode signal represents motion and the differential signal is used to detect heart rate. However, it is best to use a real motion sensor. Not only does the sensor accurately measure motion applied to the wearable device, but it can also be used to provide other functions such as tracking activity, counting steps, or launching an application when a specific g value is detected.

The ADXL362 is a micropower, 3-axis MEMS (Micro Electro Mechanical Systems) accelerometer that is ideal for detecting motion in battery-powered wearable applications. The built-in 12-bit ADC converts the acceleration value to a digital signal with a resolution of 1 mg. Power consumption varies dynamically with sample rate, consuming only 1.8μA at an output data rate of 100 Hz and 3.0 μA at 400 Hz. These higher data rates are very useful for user interfaces, such as click/double click detection.

For applications that start an application when motion is detected, high-speed sampling is not required, so the data rate can be reduced to 6 Hz with an average power consumption of 300 nA. Thus, this sensor is very attractive for low power applications and implantable devices that do not easily replace the battery. The ADXL362 is available in a 3.0 mm &TImes; 3.25 mm package. Figure 3 shows the relationship between supply current and output data rate for different supply voltage conditions.

Health monitoring of wearable devices

Figure 3. ADXL362 supply current vs. output data rate

At the heart of the system is the mixed-signal on-chip meter ADuCM350, which is connected to all of these sensors and is responsible for running the necessary software and storing, displaying or transmitting the results. The device integrates a high-performance analog front end (AFE) and a 16 MHz ARM Cortex-M3 processor core, as shown in Figure 4. The combination of AFE's flexibility and microprocessor-rich features make this chip ideal for portable and wearable applications. The configurable AFE supports almost all sensors, and its programmable waveform generator can power analog sensors using either AC or DC signals. The high-performance receive signal chain conditions the sensor signals and digitizes them using a 16-bit 160 kSPS ADC without a missing code. Among them, the integral nonlinearity (INL) / differential nonlinearity (DNL) of the latter is ±1-LSB. The receive signal chain supports any type of input signal, including voltage, current, constant potential, photocurrent, and complex impedance.

Health monitoring of wearable devices

Figure 4. Cortex-M3 with integrated AFE

The AFE operates in stand-alone mode without the intervention of the Cortex-M3 processor. The programmable timing controller controls the measurement engine and the measurement results are stored in memory via DMA. Before starting the measurement, a calibration procedure can be performed to correct the offset and drift errors in the transmit and receive signal chains. For complex impedance measurements such as blood glucose, body mass index (BMI) or tissue identification applications, the built-in DSP accelerator enables 2048 point single-frequency discrete Fourier transform (DFT) without the need for M3 processor intervention. These high-performance AFE features give the ADuCM350 an unparalleled unique advantage of other integrated solutions.

The Cortex processor supports a variety of communication ports, including I2S, USB, MIPI, and LCD display drivers (static). In addition, it includes flash, SRAM and EEPROM, and supports five different power modes to maximize battery life.

Designed for ultra-low power sensors, the ADuCM350 is limited to low speed devices. For applications that require higher processing power, use an M3 core or a Cortex-M4 processor core operating at up to 80 MHz.

What is the power consumption?

Power consumption has always been a key factor in portable and wearable devices. The devices presented in this article are designed for high performance, small size, and low power consumption, but integrating all of the devices (including batteries) in a very small package remains a challenge. Although the new battery technology achieves a higher capacity per mm3, the battery is still larger than the electronics.

Energy harvesting reduces battery size and extends battery life. There are many energy harvesting technologies, including thermoelectric, piezoelectric, electromagnetic, and optoelectronic technologies. For wearable devices, the best use of light and heat. Sensors typically do not generate a large amount of output power, so heat per joule should be captured and used. The ADP5090 ultra-low-power boost regulator (shown in Figure 5) bridges the receiver and battery. This high efficiency switch mode power supply boosts the input voltage from as low as 100 mV to 3 V. During cold start, the minimum input voltage is 380 mV when the battery is fully discharged, but in normal operation, if the battery is not fully drained or some power remains in the supercapacitor, any input as low as 100 mV The signals can be converted to higher potentials and stored for later use.

The chip is available in a tiny 3 mm &TImes; 3 mm package and can be programmed to support a wide range of energy harvesting sensors. With a maximum quiescent current of 250 nA, it supports almost all battery technologies, from lithium-ion batteries to thin-film batteries and supercapacitors. An integrated protection circuit ensures safe operation.

Health monitoring of wearable devices

Figure 5. ADP5090 Energy Harvester

in conclusion

This article describes some of the low-power products used in wearable and personal health applications, but this fast-growing market is changing rapidly.

Kara offers a wide range of illuminated and non-illuminated Rocker Switches.Ranging from 1 to 6 poles,4VA to 30 amp,with many styles of  colors and functions,especially the switches with High-Current used very widely in the welding machines. Certifications include UL, CSA, TUV, CE, and more. Kara Rocker Switches include the KR1-Series abd KR2- Series based on different size of the panel cut-out.